Instrument Accuracy Specification Calculator

Interpret a stated percentage or count specification using an explicit single-term or additive convention.

Browser-local calculations · No login · No financial values sent

Your inputs

01 / Enter

These are editable examples, not recommended rates. Enter your own numbers.

Calculator inputs

Used in every convention to center the displayed bounds. Percentage-of-reading uses its magnitude.

All measurements use this same unit. Changing the label does not convert input values.

Select addition only when the actual manufacturer specification defines an additive limit. Unselected terms are ignored.

Illustrative value; enter the actual specification.

Your result

02 / Understand

Loading the example calculation…

Show the math

Calculations retain numerical precision. Displayed quantities are rounded to significant digits, not to cents.

What this means

A specification may refer to the magnitude of a reading, an explicitly defined full scale, a span, or display counts. This tool translates only the convention you select into a symmetric absolute limit.

Worked example

At 100 V, 0.5% of reading is 0.5 V. Adding 0.2% of an explicitly defined 200 V full scale (0.4 V) and 2 counts at 0.01 V/count (0.02 V) gives ±0.92 V, only if the manufacturer specifies that additive rule.

How the math works

Reading term = |reading|×reading%/100. Full-scale term = entered full-scale magnitude×FS%/100. Span term = |upper−lower|×span%/100. Count term = integer counts×resolution. The selected term, or explicitly selected additive sum, is the limit L; displayed bounds are indication±L.

About this estimate

No root-sum-square, asymmetric, temperature, frequency, time-since-calibration or range-switch rules are inferred. All unselected components are ignored. Zero percentage/count contributions do not require their scale, endpoints or resolution. The reading remains required to center the displayed bounds. If the instrument specification cannot be represented here, consult its documentation. These are specification bounds, not measurement uncertainty, guaranteed true-value bounds or a calibration pass/fail judgment.

Reference conventions

Keep working through it

Related calculators for the next question.

Formula v1.0.0 · Reviewed · Tool ID: instrument-accuracy-specification